Automatic Glove Integrity Tester

Model GIT A4

GLOVE INTEGRITY TEST UNIT consists of a basic unit and is able to test FOUR ports at the same time. The unit to test consistency of glove, cuff ring of the glove, sleeve and cuff ring of the sleeve. Therefore all parts and interfaces of the glove port system can be tested in place. The glove integrity test unit is mobile on wheels and can be used on different Isolators. The unit is made of Aluminium/SS profile, the surface is grounded. Whereas the glove PORT CHANGE PARTS is made of Derlin.

Automatic Glove Integrity Tester
Automatic Glove Integrity Tester

Technical Specifications

Test pressure 500 pa (1000 pa initial acceptance test as per ISO-14644-7)
Stabilization pressure 500 - 750 pa (20% more than leak test pressure)
Inflation time Approx. Approx. 60 sec.
Pressure stabilization time As per guidelines
Leak test time As per guidelines
Air Pressure 4 to 6 bar
Power supply 230 V / 50 - 60 Hz