Automatic Glove Integrity Tester
Model GIT A4
GLOVE INTEGRITY TEST UNIT consists of a basic unit and is able to test FOUR ports at the same time. The unit to test consistency of glove, cuff ring of the glove, sleeve and cuff ring of the sleeve. Therefore all parts and interfaces of the glove port system can be tested in place. The glove integrity test unit is mobile on wheels and can be used on different Isolators. The unit is made of Aluminium/SS profile, the surface is grounded. Whereas the glove PORT CHANGE PARTS is made of Derlin.
- The most advanced GMP compliant glove integrity system for isolators and RABs in the pharmaceutical industry.
- The system is controlled by an HMI (Color touch pad) with a touch screen.
- This system not only allow an automated test procedure and the input of batch related data, it does also monitor all data occurring during the test.
- The data can be taken over via USB port by online printer or PDF format fill
- The Glove leak tester shall provide the total printout for test for pass or fail of test.
- The data print out can be taken via USB port in PDF format
- The data print out can be taken PC via Ethernet
- 21 CFR Part 11 compliance.
Technical Specifications
Test pressure | 500 pa (1000 pa initial acceptance test as per ISO-14644-7) |
Stabilization pressure | 500 - 750 pa (20% more than leak test pressure) |
Inflation time Approx. | Approx. 60 sec. |
Pressure stabilization time | As per guidelines |
Leak test time | As per guidelines |
Air Pressure | 4 to 6 bar |
Power supply | 230 V / 50 - 60 Hz |